Background Total replacement of the knee joint is a well-established surgical procedure performed to relieve pain and to restore the functionality of a compromised knee. Nevertheless, wear is still one of the main issues affecting joint prostheses, causing loosening and, eventually, implants failure. Surface roughness of the implant has been found to influence the mechanisms for premature wear, having relevant influence on the tribological performance. Question Joint load distribution have provided valuable insight into the knee loading patterns, showing that joint contact loads differ in medial and lateral condyles. We wondered if it is possible to find any alteration between surface roughness of medial and lateral sides and if the body mass index is in any way related to the in vivo roughening of the elements. Methods For this purpose, 11explants of total knee replacement (mean follow up of 3.2 years), with identical design, were analysed acquiring the roughness on the femoral components. Results Macroscopic inspections show that lateral condyles underwent more surface damage than the medial ones. However, surface roughness was not significantly different comparing medial and lateral sides. Statistical tests outlined a significant difference in surface roughness when distinct sizes of implants, namely 3 and 4, are considered.

Tribological Performances of Total Knee Prostheses: Roughness Measurements on Medial and Lateral Compartments of Retrieved Femoral Components

Massimiliano Merola;Alessandro Ruggiero
2019-01-01

Abstract

Background Total replacement of the knee joint is a well-established surgical procedure performed to relieve pain and to restore the functionality of a compromised knee. Nevertheless, wear is still one of the main issues affecting joint prostheses, causing loosening and, eventually, implants failure. Surface roughness of the implant has been found to influence the mechanisms for premature wear, having relevant influence on the tribological performance. Question Joint load distribution have provided valuable insight into the knee loading patterns, showing that joint contact loads differ in medial and lateral condyles. We wondered if it is possible to find any alteration between surface roughness of medial and lateral sides and if the body mass index is in any way related to the in vivo roughening of the elements. Methods For this purpose, 11explants of total knee replacement (mean follow up of 3.2 years), with identical design, were analysed acquiring the roughness on the femoral components. Results Macroscopic inspections show that lateral condyles underwent more surface damage than the medial ones. However, surface roughness was not significantly different comparing medial and lateral sides. Statistical tests outlined a significant difference in surface roughness when distinct sizes of implants, namely 3 and 4, are considered.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11386/4718776
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 10
social impact